Elabscience
QuicKey Pro Human FE (Ferritin) ELISA Kit
- SKU:
- E-OSEL-H0003
- Weight:
- 0 KGS
- Shipping:
- Calculated at Checkout
Description
QuicKey Pro Series
Get more sensitive and precise results with saving at least 1-2h comparing to traditional ELISA Kits. The new developed technology in house will help to accelerate your science research in a more efficient way.
Product Details
Properties
Assay type | Sandwich-ELISA |
Format | 96T/48T/24 |
Assay time | 1.5h |
Detection range | 62.50-4000 pg/mL |
Sensitivity | 24.31 pg/mL |
Sample type &Sample volume | serum and plasma; 50μL |
Specificity | This kit recognizes FE in samples. No significant cross-reactivity or interference between FE and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are < 10%. |
Application | This ELISA kit applies to the in vitro quantitative determination of FE concentrations in serum and plasma. |
Test Principle
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human FE. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Human FE are added to the micro ELISA plate wells. Human FE in samples (or standards) combines with the coated antibody and HRP linked detection antibody specific to FE. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of Human FE in the samples is then determined by comparing the OD of the samples to the standard curve.
Kit components & Storage
An unopened kit can be stored at 2-8℃ for 6 months. After test, the unused wells and reagents should be stored according to the table.
Item | Specifications | Storage |
---|---|---|
Micro ELISA Plate(Dismountable) | 96T: 8 wells ×12 strips 48T: 8 wells ×6 strips 24T: 8 wells ×3 strips 96T*5: 5 plates, 96T |
2-8℃, 1 months |
Reference Standard | 96T: 2 vials 48T/24T: 1 vial 96T*5: 10 vials |
2-8℃, use the reconstituted standard within 24h |
Concentrated HRP Linked Detection Ab(100×) | 96T: 1 vial, 60 μL 48T/24T: 1 vial, 30 μL 96T*5: 5 vials, 60 μL |
2-8℃(Protect from light) |
Reference Standard & Sample Diluent | 96T/48T/24T: 1 vial, 20 mL 96T*5: 5 vials, 20 mL |
2-8℃ |
HRP Linked Ab Diluent | 96T/48T/24T: 1 vial, 14 mL 96T*5: 5 vials, 14 mL |
|
Concentrated Wash Buffer(25×) | 96T/48T/24T: 1 vial, 30 mL 96T*5: 5 vials, 30 mL |
|
Substrate Reagent | 96T/48T/24T: 1 vial, 10 mL 96T*5: 5 vials, 10 mL |
2-8℃(Protect from light) |
Stop Solution | 96T/48T/24T: 1 vial, 10 mL 96T*5: 5 vials, 10 mL |
2-8℃ |
Plate Sealer | 96T/48T/24T: 5 pieces 96T*5: 25 pieces |
|
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
Technical Data
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level FE were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level FE were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 183.00 | 362.63 | 1422.41 | 190.21 | 332.97 | 1312.84 |
Standard deviation | 12.04 | 15.56 | 58.60 | 10.44 | 16.62 | 69.19 |
CV (%) | 6.58 | 4.29 | 4.12 | 5.49 | 4.99 | 5.27 |
Recovery
The recovery of FE spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum(n=8) | 87-101 | 92 |
EDTA plasma(n=8) | 90-105 | 97 |
Urine(n=8) | 99 |
Additional Information
Reactivity: |
Human |
Detection Range: |
62.50~4000 pg/mL |
Sensitivity: |
24.31 pg/mL |