Elabscience
QuicKey Pro Rat ADP/Acrp30 (Adiponectin) ELISA Kit
- SKU:
- E-OSEL-R0006
- Weight:
- 0 KGS
- Shipping:
- Calculated at Checkout
Description
QuicKey Pro Series
Get more sensitive and precise results with saving at least 1-2h comparing to traditional ELISA Kits. The new developed technology in house will help to accelerate your science research in a more efficient way.
Product Details
Properties
Assay type | Sandwich-ELISA |
Format | 96T/48T/24 |
Assay time | 1.5h |
Detection range | 31.25-1000 ng/mL |
Sensitivity | 6.43 ng/mL |
Sample type &Sample volume | serum and plasma; 50μL |
Specificity | This kit recognizes ADP/Acrp30 in samples. No significant cross-reactivity or interference between ADP/Acrp30 and analogues was observed. |
Reproducibility | Both intra-CV and inter-CV are < 10%. |
Application | This ELISA kit applies to the in vitro quantitative determination of ADP/Acrp30 concentrations in serum and plasma. |
Test Principle
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Rat ADP/Acrp30. Samples (or Standards) and Horseradish Peroxidase (HRP) linked antibody specific for Rat ADP/Acrp30 are added to the micro ELISA plate wells. Rat ADP/Acrp30 in samples (or standards) combines with the coated antibody and HRP linked detection antibody specific to ADP/Acrp30. Excess conjugate and unbound sample or standard are washed from the plate. The substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450±2 nm. The concentration of Rat ADP/Acrp30 in the samples is then determined by comparing the OD of the samples to the standard curve.
Kit components & Storage
An unopened kit can be stored at 2-8℃ for 6 months. After test, the unused wells and reagents should be stored according to the table.
Item | Specifications | Storage |
---|---|---|
Micro ELISA Plate(Dismountable) | 96T: 8 wells ×12 strips 48T: 8 wells ×6 strips 24T: 8 wells ×3 strips 96T*5: 5 plates, 96T |
2-8℃, 1 months |
Reference Standard | 96T: 2 vials 48T/24T: 1 vial 96T*5: 10 vials |
2-8℃, use the reconstituted standard within 24h |
Concentrated HRP Linked Detection Ab(100×) | 96T: 1 vial, 60 μL 48T/24T: 1 vial, 30 μL 96T*5: 5 vials, 60 μL |
2-8℃(Protect from light) |
Reference Standard & Sample Diluent | 96T/48T/24T: 1 vial, 20 mL 96T*5: 5 vials, 20 mL |
2-8℃ |
HRP Linked Ab Diluent | 96T/48T/24T: 1 vial, 14 mL 96T*5: 5 vials, 14 mL |
|
Concentrated Wash Buffer(25×) | 96T/48T/24T: 1 vial, 30 mL 96T*5: 5 vials, 30 mL |
|
Substrate Reagent | 96T/48T/24T: 1 vial, 10 mL 96T*5: 5 vials, 10 mL |
2-8℃(Protect from light) |
Stop Solution | 96T/48T/24T: 1 vial, 10 mL 96T*5: 5 vials, 10 mL |
2-8℃ |
Plate Sealer | 96T/48T/24T: 5 pieces 96T*5: 25 pieces |
|
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
Technical Data
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level ADP/Acrp30 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level ADP/Acrp30 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (ng/mL) | 52.94 | 153.56 | 371.41 | 50.14 | 155.02 | 346.08 |
Standard deviation | 2.94 | 8.06 | 11.14 | 2.58 | 8.79 | 17.65 |
CV (%) | 5.55 | 5.25 | 3.00 | 5.15 | 5.67 | 5.10 |
Recovery
The recovery of ADP/Acrp30 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
---|---|---|
Serum(n=8) | 87-100 | 95 |
EDTA plasma(n=8) | 95-111 | 101 |
Additional Information
Reactivity: |
Rat |
Detection Range: |
31.25~1000 ng/mL |
Sensitivity: |
6.43 ng/mL |