Detection principle
ABTS is oxidized to green ABTS•+ by appropriate oxidant, which can be reduced to colorless ABTS in the presence of antioxidants. The TAS of the sample can be determined and calculated by measuring the absorbance of ABTS•+ at 660 nm. Trolox is an analog of VE and has a similar antioxidant state to that of VE. Trolox is used as a reference substance for total antioxidant status.
Performance characteristics
Synonyms | TAS |
Sample type | serum, plasma, urine, cellular supernatant, animal, plant tissue |
Sensitivity | 0.23 mmol Trolox Equiv./L |
Detection range | 0.23-2 mmol Trolox Equiv./L |
Detection method | Colorimetric method |
Assay type | Quantitative |
Assay time | 30 min |
Precision | Average inter-assay CV: 7%Average intra-assay CV: 4.6% |
Other instruments required | Micropipettor, 37℃ incubator. |
Other reagents required | 60% Ethanol |
Storage | -20℃ |
Valid period | 6 months |
Images
TAS significantly decreased in diabetic model (DM), treatment of Cellgevity® significantly improved TAS. (**P<0.01, ***P<0.001)
O O Ogunlabi et al investigate the effects of Cellgevity® in diabetic model. Total antioxidants status (TAS) of rat serum was determined using TAS colorimetric assay kit (E-BC-K801-M).
Dilution of sample
It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.23-2 mmol Trolox Equiv. /L).
The recommended dilution factor for different samples is as follows (for reference only):
Sample type |
Dilution factor |
10% Mouse liver tissue homogenate |
1 |
10% Rat liver tissue homogenate |
1 |
10% Rat lung tissue homogenate |
1 |
Molt4 cellar supernatant |
1 |
Human urine |
8-10 |
Mouse serum |
1 |
Human serum |
1 |
Human saliva |
1 |
Note: The diluent is 60% Ethanol.
Additional Information
Detection instrument: |
Microplate reader(650 nm-670 nm,optimum wavelength: 660 nm) |
Detection method: |
Colorimetric method |